Home  |  About Us  |  Products  |  Certificates  |  Events  |  Feed Back  |  Contact Us  |  Email  |  Site Map
 
 
RESEARCH AIDS
     
 
  APEL SPECTROPHOTOMETER
418VL PD-303 APEL SPECTROPHOTOMETER
418VL PD-303
 SMALL SAMPLE VOLUME
Minimum sample volume is 1.0ml
 LONG LAMP LIFE
Long life span of lamp (approx. 2000 hours)
 SQUARE, ROUND CUVETTE USABLE
Square and round cuvette can be used without adapter
 FLEXIBLE POWER SOURCE
Power requirment is flexible from100 to 240VAC
 ANALOG OUTPUT
Possible to output the results to the recorder using supplied analogue output cable
   Display
  3½ Digit LCD
   Wavelength Range
  340 - 1000 nm
   Wavelength Width
  10nm in entire region
   Wavelength Accuracy
  ± 2 nm at 365 and 546nm
   Wavelength Scale
  2nm scale
   
  APEL SPECTROPHOTOMETER
418VL PD-303S APEL SPECTROPHOTOMETER
419VL PD-303S
 PARAMETERS
5-parameters type is also available. (15 Standard curves can be memorized) current model is 5-parameters type.
 EXTERNAL INTERFACE
Printer interface and RS-232C are built in.
 AUTO-ZERO FUNCTION
T=0% and T=100% can be adjusted with one touch of a button
 SMALL SAMPLE
Minimum sample volume is 1.0 ml.
 ROUND OR SQUARE CUVETTE USABLE
Either round cuvette or square cuvette can be used without adapter.
 MEMORY AND OTHER FUNCTIONS
Microprocessor enables various functions such as the measurement of correlation curve, timer mode and memory of standard data etc. as standard functions
   Display
  LCD Display
   Wavelength Range
  340 - 1000 nm
   Wavelength Width
  8nm in entire region
   Wavelength Accuracy
  ± 2 nm at 430 and 585 nm
   Wavelength Scale
  1nm scale
   
  APEL SPECTROPHOTOMETER
418VL PD-303UV APEL SPECTROPHOTOMETER
419VL PD-303UV
   Display
  2-lines, 20 character LCD Display
   Wavelength Range
  190 - 1000nm
   Wavelength Width
  5nm
   Wavelength Accuracy
  ± 2 nm
   Wavelength Scale
  1nm scale (Readable)
   Wavelength Repeatability
  ± 0.5 nm
   Optical System
  Single beam, grating system 1200 lines / mm
   Stray Radiant Energy
  <_0.3%T typical at 220nm and 340nm
   Photometric Range
  0 - 2.5 ABS / 0 - 125.0% T / 0 - 1999 C (0-1999F)
   Photometric Accuracy
  ± 0.004A at 0.5A
   Drift
  Less than 0.002A/hour after warm up
   Detector
  Silicon Photocell
   Light Source
  Tungten halogenlamp, deuterium lamp
   Auto Zero
  Yes
   Standard Sample Holder
  4 Position, 10x10mm squre cuvette
   Data Output
  Analog output, RS232C (D-sub 9pins)
   Application Software
  Included
   


 
All right reserved. © www.axivasichem.com